55
Q-TECH™ ProgrammEr: oPEraTIon
Manual Setup
Manual Setup enables the user to perform the electrode integrity test and select the electrode
sensing configuration and gain setting in the SQ-RX device.
1.
From the Utilities Menu, select the Manual Setup button. The Impedance Test
screen appears (Figure 45).
2.
Select the Test button to perform the electrode integrity test.
Figure 45:
Impedance test screen
3.
Select Continue to proceed or select Test to retry. The sensing configuration can be
evaluated by temporarily activating any of the vectors or gain settings. The Live S-ECG
and markers will reflect the temporary selections for evaluation purposes.
4.
There are three available sensing vectors that can be manually selected:
•
primary:
Sensing from the proximal electrode ring on the subcutaneous electrode
to the surface of the active SQ-RX device
•
Secondary:
Sensing from the distal sensing electrode ring on the subcutaneous
electrode to the surface of the active SQ-RX device
•
Alternate:
Sensing from the distal sensing electrode ring to the proximal sensing
electrode ring on the subcutaneous electrode