13. Connect the ESU-2400 to the DUT. Connect the Yellow cable from the Active port
on the ESU-2400 to active output on the DUT. Connect the Blue cable from the
Dispersive ports of the ESU-2400 to the return on the DUT.
14. Press Save Configuration on the ESU-2400 to save the Load Curve settings to a
file for future use, or Press Start Load Curve to begin the test. You will be prompted
when to start and stop triggering the generator.