72
Step Values Tab
– This tab allows the user to configure a series of CQM resistances to
be stepped through. The CQM resistance can be configured to step by a fixed resistance
amount or by a fixed number of steps across a range or resistances. At each step, the
DUT CQM state is evaluated. The test can be configured to stop when either an alarm
or normal CQM state occurs.
The Set Tolerance button is used to determine the pass / fail parameters. The CQM
resistance will be stepped until the CQM state is either normal or alarm as programmed
in the step. This point is called the trip resistance. The tolerance configuration allows the
user to configure the trip points that result in a pass or fail status for the test.