94
In the above illustration the red line represents the user-selectable threshold 1. The yellow
line represents the user-selectable threshold 2. The purple signal represents when the
waveform is above threshold 1 and the ESU-
2400H is measuring it as “pulse 1”. The blue
signal represents when the waveform is below threshold 1 but above threshold 2 and is
being measured as “pulse 2”. The bottom green signal represents when the waveform is
below threshold 2 and is being measured as “pulse 3”.
The above image shows typical measurements when “All Readings” is selected from
“Display Parameters”. The voltage, current, power, and time for each “pulse” is displayed,
as well as the duty cycle for all three pulses.