The following table shows all of the faults that are monitored by the system.
Message
Description
Recovery Action
Left Fan
Blocked
The left fan is not spinning.
Permanent damage or complete
failure could occur.
Check the fan for obstructions. If
nothing is found and the fan is not
spinning contact BC Biomedical
for service.
Right Fan
Blocked
The right fan is not spinning.
Permanent damage or complete
failure could occur.
Check the fan for obstructions. If
nothing is found and the fan is not
spinning contact BC Biomedical
for service.
System
Overtemp
The internal temperature has
exceeded normal operating
conditions.
Stop activating the DUT, allow the
system to cool for at least a half
hour. If the problem persists,
contact BC Biomedical for
service.
Footswitch
Fuse Open
The internal fuse protecting the
power supply to the footswitch
interface has opened.
Contact BC Biomedical for
service.
Front USB
Fuse Open
The internal fuse protecting the
power supply to the front USB port
has opened.
Contact BC Biomedical for
service.
Side USB
Fuse (Front)
Open
The internal fuse protecting the
power supply to the side USB port
towards the front of the unit has
opened.
Contact BC Biomedical for
service.
Side USB
Fuse (Rear)
Open
The internal fuse protecting the
power supply to the side USB port
towards the rear of the unit has
opened.
Contact BC Biomedical for
service.
No Response
from DUT
The ESU-2400 expected RS-232
data from the DUT and did not
receive it.
Check the RS-232 cable to
ensure that it is seated completely
on the ESU-2400 and DUT.
ForceTriad
Version Error
The Autosequences are only
compatible with ForceTriads running
firmware 3.50, 3.60, 3.80, or 4.00.
This error will occur if you are
trying to use a ForceTriad PM
Autosequence on a unit with
software version less than 3.50.
Go to
www.valleylab.com
to get a
software update for your
ForceTriad.
Summary of Contents for ESU-2400
Page 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Page 2: ......
Page 9: ...vii...
Page 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 177: ...165 This Page Intentionally Left Blank...
Page 186: ...174 NOTES...
Page 187: ...175 NOTES...
Page 188: ...176 NOTES...
Page 189: ......