73
List Values Tab
– This allows the user to configure a list of CQM resistances that are
used in the test. During the test the CQM resistance will be sequentially set to the values
shown in the CQM resistance list. The step can be configured to stop when the CQM
state is either alarm or normal.
The Set Tolerance button is used to determine the pass / fail parameters. The CQM
resistance will be stepped until the CQM state is either normal or alarm as programmed
in the step. This point is called the trip resistance. The tolerance configuration allows the
user to configure the trip points that result in a pass or fail status for the test.