DELAYED MODE
Delayed Mode is a special form of Continuous Mode that is used for signals that have a
long settling time or other portion at the beginning of the activation that the user does not
wish to measure, such as noise.
The green signal represents a waveform to be measured in Delayed Mode.
The red line represents the user-selectable threshold.
The purple signal represents when the waveform has crossed the threshold and has been
detected as a the beginning of the RF output from the generator
The blue signal represents the start of the measurement. Rather than measuring upon
detecting the pulse it has been delayed using a user-selectable delay value.