8. Press the Load resistance setting for the First Load and enter 50 in the number
pad.
9. Press the Load resistance setting for the Last Load and enter 5000 in the number
pad.
10. Press the step setting in the Total Steps in Test option and enter 10 in the number
pad. The screen should now look like the following:
11. Press the Setup Power tab at the top of the screen to configure the power setting
for the test.
12. Select the Single Power Level tab to configure the test for one power setting.
13. Press the DUT Power setting button and enter 300 in the number pad. The screen
should look like the following:
14. The configuration is complete for this step. Press Back to Step Edit and then Done
to return to the Autosequence Summary screen.