48
7. Press the Test Load button to open the Load Configuration menu.
8. Press the Internal Load value button to change the internal load to 200 ohms. By
default, the user will not be able to change the load resistance once the test is
running. If you would like for the operator to be able to change the load, check the
box next to Allow Operator to modify load.
9. Press the Test Load button again or the Red X button to close the Load
Configuration menu.
10. Press the Tolerance option button at the bottom of the screen to configure the
tolerance for the RF measurement.
11. The DUT output mode and output type and power setting can be selected at the
top of the screen. These are informational only and do not have any effect on the
test or measurement.
12. Use the dropdown arrow under Expected Reading to change the tolerance mode
to Equal to (Value).
13. Press the number button to the right of the expected reading and enter 315 in the
number pad. Additionally, if the desired reading were watts or any other
measurement taken by the ESU-2400, the red button to the right of the expected
reading can be pressed to change the measurement being tested.
14. Press the number button for High Limit and set the value to 25 with the number
pad.
15. Press the number button for Low limit and set the value to 25 with the number pad.
Summary of Contents for ESU-2400
Page 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Page 2: ......
Page 9: ...vii...
Page 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 177: ...165 This Page Intentionally Left Blank...
Page 186: ...174 NOTES...
Page 187: ...175 NOTES...
Page 188: ...176 NOTES...
Page 189: ......