35
Leakage Test 2
– I need to measure the leakage current from the dispersive (REM) port
to ground. I need to have a 200 ohm load from Active to Dispersive during the test.
Solution
–
1. Select Measure RF Leakage from the main menu.
2. If the Leakage selection at the top of the screen does not show Leakage 2, press
the button and select Leakage 2 from the test mode menu.
3. Connect the Active output from the DUT to the Active input on the ESU-2400. If
you are using the cables from the ESU-2400 accessory kit, use the Yellow Lead.
4. Connect the Dispersive output from the DUT to the Dispersive inputs on the
ESU-2400. If you are using the cables from the ESU-2400 accessory kit, use the
Blue cable without the clear pin.
5. Connect earth ground port on the ESU-2400 to the ground lug on the back of the
DUT. If you are using the cables from the ESU-2400 accessory kit, use the
green cable and green alligator clip.
NOTE: To see how the ESU-2400 is internally configured, press the Test Setup button. To close
the image, press Test Setup again.
Summary of Contents for ESU-2400
Page 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Page 2: ......
Page 9: ...vii...
Page 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 177: ...165 This Page Intentionally Left Blank...
Page 186: ...174 NOTES...
Page 187: ...175 NOTES...
Page 188: ...176 NOTES...
Page 189: ......