
150
ESU-2400 Communication Command Summary (Continued)
Keywords
Nodes
Subnodes
Values
CONFigure
CQM
LOAD
OPEN or 0-500 (ohms)
Set or read CQM
resistance
COHMs
0-500 (ohms)
Set or read Change by
ohms amount
UP
Same as pressing change
by ohms UP button
DOWN
Same as pressing change
by ohms DOWN button
POHMs
0-500 (percent)
Set or read Change by
Percent amount
UP
Same as pressing change
by Percent UP button
DOWN
Same as pressing change
by Percent DOWN button
MODE
MAIN, RFMeasure, RFLeakage,CQM,Lcurve,ASEQuence, SYStools
HOLD
ON,OFF (only for RFMeasure, RFLeakage modes)
SYSTem
VERsion?
[read only]
VOLume
0-10
AVERaging
FAST
1-200 (samples)
MEDium
1-200 (samples)
SLOW
1-200 (samples)
WINdow
0.0 - 100.0 (mV)
READ
MVrms?
Returns: mV RMS [read only]
MArms?
Returns: mA RMS [read only]
WArms?
Returns: Watts RMS [read only]
MVPeak?
Returns: mV Peak [read only]
MVPP?
Returns: mV Peak to Peak [read only]
MVP-PP?
Returns: mV Peak/Peak to Peak [read only]
MVPK+?
Returns: mV Positive Peak [read only]
CF?
Returns: Crest Factor [read only]
KHZ?
Returns: Frequency [read only]
TON?
Returns: Time - Pulse On [read only]
TOFF?
Returns: Time - Pulse Off [read only]
TCYC?
Returns: Time - Total Cycle [read only]
DCYC?
Returns: % Duty Cycle [read only]
MVCyc?
Returns: mV Total Cycle [read only]
MACyc?
Returns: mA Total Cycle [read only]
WCyc?
Returns: Watts Total Cycle [read only]
ALL?
Returns: All readings in comma delimited format
DATA?
Returns: Entire Data Buffer [read only] Length = NUMSamples * 2
Data Format is mV in signed double byte format, Decimal place is assumed based
on Input Range (Low Range = 2DP, High Range = 1DP)
TEMP?
Returns system temperature (Highest temperature of 8 sensors)
SVOLtage?
Returns List of system supply voltages
Footswitch Fuse, +12V, Vbacklight, +5V, +3D, +3ETX, +3DISP, VPS2, Front USB,
Side USB Bottom, Side USB Top
Summary of Contents for ESU-2400
Page 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Page 2: ......
Page 9: ...vii...
Page 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 177: ...165 This Page Intentionally Left Blank...
Page 186: ...174 NOTES...
Page 187: ...175 NOTES...
Page 188: ...176 NOTES...
Page 189: ......