Scenario 4 (ESU-2400H Model Only)
– I need to measure the frequency and mA
output of a generator with a pulsed waveform using a 200 ohm load. I know that the
waveform has a 700 mA pulse but little or no current between pulses.
I don’t have a
footswitch cable so I will trigger the DUT myself.
Solution
–
1. Select Measure RF Energy from the main menu.
2. Connect the Active output from the DUT to the Active input on the ESU-2400. If
you are using the cables from the ESU-2400 accessory kit, use the Yellow Lead.
3. Connect the Dispersive or Return port on the DUT to the Dispersive 1 and
Dispersive 2 inputs on the ESU-2400. If you are using the cables from the ESU-
2400 accessory kit, use the Blue cable without the pin.
4. Press the Isolated Generator Type button to configure the ESU-2400 for isolated
generator measurements.
5. Press the Input M
ode button, which by default reads “Continuous Waveform”,
and select Pulsed from the dialog options.