Auto CQM
– I need to test the CQM input on my generator. My generator service manual
says to increase the CQM resistance starting at 120 ohms until the CQM alarm occurs.
The CQM trip resistance should be 135 ohms +/- 5 ohms.
Solution
–
1. From the Autosequence menu, select Add Step.
2. Select the desired location for the new step and press OK.
3. The new step will always default to an instruction type step. Select the
REM/ARM/CQM option button near the bottom of the screen.
4. Press the Title text box to change the step title as desired. You will need to use
an external keyboard for this.
5. Press the Instructions to Operator text box to enter any instructions that the user
might need to perform at this step of the autosequence.
6. Press the Meter Setup button to configure the settings for the REM test.
7. Select the Auto CQM option at the top of the screen.