33
Leakage Test 1b
– I need to measure the leakage current from the dispersive (REM)
port to ground.
Solution
–
1. Select Measure RF Leakage from the main menu.
2. If the Leakage selection at the top of the screen does not show Leakage 1b,
press the button and select Leakage 1b from the test mode menu.
3. Connect the Dispersive output from the DUT to the Dispersive inputs on the
ESU-2400. If you are using the cables from the ESU-2400 accessory kit, use the
Blue cable without the clear pin.
4. Connect earth ground port on the ESU-2400 to the ground lug on the back of the
DUT. If you are using the cables from the ESU-2400 accessory kit, use the
green cable and green alligator clip.
NOTE: To see how the ESU-2400 is internally configured, press the Test Setup button. To close
the image, press Test Setup again.