43
14. Use the dropdown arrow to set the DUT mode to the desired bipolar output mode.
The screen should now look like the following:
15. Connect the ESU-2400 to the DUT. Connect the Yellow cable from the Active port
on the ESU-2400 to one of the Bipolar outputs on the DUT. Connect the Green
cable from the Dispersive 1 port of the ESU-2400 to the other Bipolar output on
the DUT.
16. Connect the RS-232 cable from the ESU-2400 serial port to the DUT serial port.
17. Connect the BC20-03001 footswitch cable from the DUT to the ESU-2400.
18. Press Save configuration on the ESU-2400 to save the Load Curve settings to a
file for future use, or Press Start Load Curve to begin the test. The DUT power
level and output mode will automatically be configured through the RS-232 port.