
28
Scenario 5
(ESU-2400H Model Only)
– I need to measure the load voltage of a
generator that has a pulsed waveform with a long settling time of about 2 seconds. I
don’t have a footswitch cable so I will trigger the DUT myself.
Solution
–
1. Select Measure RF Energy from the main menu.
2. Connect the Active output from the DUT to the Active input on the ESU-2400. If
you are using the cables from the ESU-2400 accessory kit, use the Yellow Lead.
3. Connect the Dispersive or Return port on the DUT to the Dispersive 1 and
Dispersive 2 inputs on the ESU-2400. If you are using the cables from the ESU-
2400 accessory kit, use the Blue cable without the pin.
4. Press the Isolated Generator Type button to configure the ESU-2400 for isolated
generator measurements.
5.
Press the Input Mode button, which by default reads “Continuous Waveform”,
and select Delayed from the dialog options.
NOTE
: Delayed Mode is designed for use with pulsed waveforms only.
Summary of Contents for ESU-2400
Page 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Page 2: ......
Page 9: ...vii...
Page 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 177: ...165 This Page Intentionally Left Blank...
Page 186: ...174 NOTES...
Page 187: ...175 NOTES...
Page 188: ...176 NOTES...
Page 189: ......