160
ESU-2400H Communication Command Summary (Continued)
Keywords
Nodes
Subnodes
Values
SYSTem
VERsion?
[read only]
VOLume
0-10
AVERaging
FAST
1-200 (samples)
MEDium
1-200 (samples)
SLOW
1-200 (samples)
WINdow
0.0 - 100.0 (mV)
READ
VLOAD?
Returns: Load Voltage [read only]
MA?
Returns: mA RMS [read only]
WATTS?
Returns: Watts RMS [read only]
VPeak?
Returns: V Peak [read only]
VPP?
Returns: V Peak to Peak [read only]
VP-PP?
Returns: V Peak/Peak to Peak [read only]
VPK+?
Returns: V Positive Peak [read only]
CF?
Returns: Crest Factor [read only]
KHZ?
Returns: Frequency [read only]
TP1?
Returns: Time - Pulse 1 On [read only]
TOFF?
Returns: Time - Pulse Off in Pulsed Mode 1 [read only]
TCYC?
Returns: Time - Total Cycle [read only]
DP1?
Returns: % Duty Cycle of Pulse 1 [read only]
VP1?
Returns: Load Voltage of Pulse 1 [read only]
MAP1?
Returns: mA Current of Pulse 1 [read only]
WP1?
Returns: Watts Pulse 1 [read only]
VP2?
Returns: Load Voltage of Pulse 2 [read only]
MAP2?
Returns: mA Current of Pulse 2 [read only]
WP2?
Returns: Watts Pulse 2 [read only]
VPKP2?
Returns: Peak Voltage of Pulse 2 [read only]
TP2?
Returns: time - Pulse 2 On [read only]
DP2?
Returns: % Duty Cycle of Pulse 2 [read only]
VP3?
Returns: Load Voltage of Pulse 3 [read only]
MAP3?
Returns: mA Current of Pulse 3 [read only]
WP3?
Returns: Watts Pulse 3 [read only]
VPKP3?
Returns: Peak Voltage of Pulse 3 [read only]
TP3?
Returns: time - Pulse 3 On [read only]
DP3?
Returns: % Duty Cycle of Pulse 3 [read only]
ALL?
Returns: All readings in comma delimited format
TEMP?
Returns system temperature (Highest temperature of 8 sensors)
SVOLtage?
Returns List of system supply voltages
Footswitch Fuse, +12V, Vbacklight, +5V, +3D, +3ETX, +3DISP, VPS2, Front USB,
Side USB Bottom, Side USB Top
Summary of Contents for ESU-2400
Page 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Page 2: ......
Page 9: ...vii...
Page 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 177: ...165 This Page Intentionally Left Blank...
Page 186: ...174 NOTES...
Page 187: ...175 NOTES...
Page 188: ...176 NOTES...
Page 189: ......