143
ESU-2400 Model Only
CONFigure Subsystem
This group allows the user to setup the display and operational settings for the unit.
Note that there are independent settings for the Measure RF Energy screen and the
Measure RF Leakage screen.
KEYWORD
PARAMETER FORM
COMMENTS
CONFigure
:RFMeasure
:DISPlay
:SxZy
nn
S = display_screen_number
Z = zone_number
display_screen_number = 1-6
nn = Parameter for selected Zone
1 = One Parameter
0 = mV RMS
8 = kHz
2 = Two Parameters
1 = mA RMS
9 = Time-Pulse On
3 = Three Parameters
2 = Watts RMS
10 = Time-Pulse Off
4 = Four Parameters
3 = mV Peak
11 = Time-Total Cycle
5 = Five Parameters
4 = mV Pk-Pk
12 = % Duty Cycle
6 = Display All Parameters
5 = mV Pk / mV Pk-Pk
13 = mV Pulse Cycle
6 = mV Pk+
14 = mA Pulse Cycle
7 = Crest Factor
15 = Watts Pulse Cycle
:SCReen < numeric_value >
:AVERaging FAST | SLOW | MEDium
:LOAD
:MODE
INTernal | EXTernal | INT/EXTernal
:INTernal <numeric_value>
0-6400
:EXTernal <numeric_value>
0-6400
:ACTual?
Read Only, returns calibrated
load value, what the 2400 uses
for Watts calculations
:INPut
:RANGe 100 | 1000 | AUTo
:MODE
CONTinuous | PULsed
:TYPE
ISOlated | GND
Range 1-6
1-5 = # display zones
6=Measurement List Display
Summary of Contents for ESU-2400
Page 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Page 2: ......
Page 9: ...vii...
Page 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 177: ...165 This Page Intentionally Left Blank...
Page 186: ...174 NOTES...
Page 187: ...175 NOTES...
Page 188: ...176 NOTES...
Page 189: ......