39
Scenario 1
– I need to test the REM function on my generator. The service manual
says that the REM alarm should be Normal at 35 ohms.
Solution
–
1. Select REM/ARM/CQM from the main menu.
2. Press the CQM Resistance button and enter 35 in the number pad. The screen
should look like the following:
3. Connect the Dispersive output from the DUT to the Dispersive inputs on the
ESU-2400. If you are using the cables from the ESU-2400 accessory kit, use the
Blue cable with the center pin.
4. The DUT REM state should be Normal.
REM/ARM/CQM TEST