14
Leakage 1b:
This leakage test, specified by the IEC as Dispersive Electrode to Ground,
is for testing the RF leakage to earth ground of an Isolated Output type CF electrosurgical
generator from a single active or dispersive lead. The test complies with IEC 601.2.2,
sec. 19.101b, fig, 104 and sec. 19.102, adopted by ANSI/AAMI HF18-2001. The purpose
of this test is to verify that open circuit RF leakage of the DUT meets or exceeds the IEC
specification. The ESU-2400 internally connects the load to earth ground. An external
ground is also provided at the front panel.
NOTE:
DO NOT TEST AN EARTH REFERENCED TYPE BF ELECTROSURGICAL
GENERATOR WITH THIS TEST, THE RESULTING MEASUREMENT WILL BE
ERRONEOUS.
Connections
1. Dispersive 1 Jack on ESU-2400 to DUT Dispersive Output or Bipolar 2.
WARNING
– ONE LEAD AT A TIME
Only test one lead of the ESU Generator at a
time, either Active or Dispersive, not both.
Earth Ground to
DUT equipotential
jack
ESU (DUT)
Dispersive
Electrode
Summary of Contents for ESU-2400
Page 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Page 2: ......
Page 9: ...vii...
Page 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 177: ...165 This Page Intentionally Left Blank...
Page 186: ...174 NOTES...
Page 187: ...175 NOTES...
Page 188: ...176 NOTES...
Page 189: ......