4. Press the Test Load button to open the load configuration box. The screen
should look like the following:
NOTE: The ESU-2400 can be configured to use internal loads, external loads, or a combination of
both. First select the Load Mode, then enter the desired internal and external loads. The total load
will automatically be updated. For most testing, the internal loads will be sufficient.
5. Press the Button next to Internal Load and enter 100 in the number pad. The
screen should automatically return to the Measure RF screen.
6. Trigger the DUT. The mA measurement will be shown on the ESU-2400 Screen.
NOTE: The displayed measurement can be changed by pressing the mA RMS button. There are
many measurements that can be displayed.