7. Press the Configure Footswitch button to open the Footswitch Setup box.
8. Use the dropdown arrow under Select Output to select the Bipolar output.
9. Use the dropdown arrow under Activation Timeout to select the maximum time
the DUT will be triggered.
NOTE: The footswitch output selection is only valid for footswitch cables that are designed for a
specific DUT, such as the BC20-03001 or BC20-03002. If you are using the unterminated
footswitch cable, BC20-03000, the output selection is dependent on how the cable is wired.
10. Press the footswitch button to trigger the DUT. The mA and Watts
measurements will be shown on the ESU-2400 Screen.
11. You may either wait for the Footswitch Activation Timeout, or press the
Footswitch button again to stop triggering the DUT output.