155
READ Subsystem
This group allows the user to get measurements from the unit.
KEYWORD
PARAMETER FORM
COMMENTS
READ:
V Load | MA | WATTS | VPeak |
Read Only
VPP | CF | KHZ
TP1 | TOFF | TCYC | DP1
Read Only (Pulsed Mode 1)
VP1 | MAP1 | WP1
VP2 | MAP2 | WP2 | VPKP2
Read Only (Pulsed Mode 2)
TP2 | DP2
VP3 | MAP3 | WP3 | VPKP3
Read Only (Pulsed Mode 3)
TP3 | DP3
ALL?
Returns All readings in comma
delimited format
TEMP?
Returns system temperature
SVOLtage?
Returns comma delimited list of
system supply voltages:
Footswitch Fuse, +12V,
Vbacklight, +5V, +3D, +3ETX,
+3DISP, VPS2, Front USB,
Bottom USB, Top USB
STATus Sub-system
This subsystem provides status on the operating mode of the unit including messages
that would normally be seen on the display.
KEYWORD
PARAMETER FORM
COMMENTS
STATus?
Read Only
Bit Value
Definition
0
1
Hold Mode
1
2
2
4
Calibration Mode
3
8
4
16
Tissue Response Mode
5
32
6
64
Tissue Response Test Complete
7
128
Tissue Response Test Started
8
256
Error Present
9
512
mV Out of Range
10
1024
mA Out of Range
11
2048
Watts Out of Range
12
4096
mV Peak Out of Range
13
8192
Crest Factor Out of Range
14 16384
Fan 1 Blocked
15 32768
Fan 2 Blocked
Summary of Contents for ESU-2400
Page 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Page 2: ......
Page 9: ...vii...
Page 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Page 177: ...165 This Page Intentionally Left Blank...
Page 186: ...174 NOTES...
Page 187: ...175 NOTES...
Page 188: ...176 NOTES...
Page 189: ......