Leakage 2
: This leakage test, specified by the IEC as Earth Reference Leakage Type BF
(Load Between Electrodes), is for testing the leakage to earth ground of a Ground
Referenced Output type BF electrosurgical generator from the active output. This test
complies with IEC 601.2.2, sec. 19.101a, test 1, fig. 102, and sec. 19.102, adopted by
ANSI/AAMI HF18-2001. The purpose of this test is to verify that the RF leakage of the
DUT meets or exceeds the IEC specification.
Connections
1. Active Lead from Active Jack on ESU-2400 to DUT Active Output.
2. Dispersive Lead from Dispersive Jack on ESU-2400 to DUT Dispersive.
ESU (DUT)
Active
Electrode
ESU (DUT)
Patient Return/
Dispersive
Electrode