T405S Indicator Service Manual
7
Communication port protocols
<LF>XC<CR>
This will return the audit counters in this format:
<LF>Calib:xxx:Config:yyy:<CR>
<LF>XD<n><CR>
This will initiate the resetting of the calibration and/or
configuration parameters depending on the format
chosen from the list below:
If n=1, reset config only
If n=2, reset calibration only
If n=3, reset both
The response will be in the form of:
<LF>xd<n><CR>
<LF>XZ<CR>
This will initiate a Cal Zero command. The scale
responds with a standard scale response message,
with one additional custom status response added. If
the “Set Zero” operation fails then the scale responds
with the <s> field = “0”
<LF>XS<wwww><CR>
This will initiate a Cal Span command where <wwww>
is the weight used to calibrate with. The scale responds
with a standard scale response message, with one
additional custom status response added. If the “Set
Span” operation fails then the scale responds with the
<s> field = “s”
<LF>XVS<n>:<value><CR>
The indicator will set the <value> of the variable
specified by the Network token. A valid XVS command
will receive a response of <LF>xvs<CR>. To verify an
XVS command was successful requires sending the
XVG command to confirm the value of the variable.
<LF>XVG<n><CR>
The indicator will get or return the value of the variable
specified by the network token. A valid response is in
the form of <LF><value><CR>.
Summary of Contents for T405S
Page 1: ...AWT35 501638 Issue AA T405S Weight Indicator Service Manual...
Page 2: ...BTEK_T405S_s_en_501638 book...
Page 8: ...8 T405S Indicator Service Manual...
Page 134: ...134 T405S Indicator Service Manual...
Page 147: ...T405S Indicator Service Manual 147 13 7 Outline dimensions...
Page 150: ...150 T405S Indicator Service Manual...
Page 154: ...154 T405S Indicator Service Manual...
Page 155: ......
Page 156: ...B TEK Scale LLC 1510 Metric Ave SW Canton OH 44706 3088...