T405S Indicator Service Manual
75
6.5
Ports
6.5.7 File
FiLE
Use this item to configure how and where files are saved.
PAth
Use this to choose where the file is saved, internally for
transfer by
FtP
using an Ethernet port or to a connected
uSb
flash drive.
ModE
Use this to configure the file convention. Choices are:
conStAn
This will create a single file that will be appended
to as new information is saved.
nuMbrd
This will create new file appended with a
sequential number for each transaction.
D_T
File
Path
Mode
Reset
USB
FTP
Numbrd
Constan
Resets
to 000001
When the FILE > PATH is set for USB, a USB flash drive can be installed into the USB
Host port and transactions will be transferred to the flash drive as they occur. If the
flash drive is not installed, transactions will be saved onto the Flash memory in the
indicator.
The size of the print format will determine the number that can be stored but with a
typical print format the indicator should be capable of storing several thousand
transactions. The saved transactions will transfer immediately when the USB flash
drive is re-installed. The message
CoPYing
will display as each stored file is
transferred to the flash drive. If using the d_t file mode this could take a few seconds
to transfer each file. Do not remove the flash drive until the message
donE
is finally
displayed. Stored transactions will then be cleared after the transfer.
Summary of Contents for T405S
Page 1: ...AWT35 501638 Issue AA T405S Weight Indicator Service Manual...
Page 2: ...BTEK_T405S_s_en_501638 book...
Page 8: ...8 T405S Indicator Service Manual...
Page 134: ...134 T405S Indicator Service Manual...
Page 147: ...T405S Indicator Service Manual 147 13 7 Outline dimensions...
Page 150: ...150 T405S Indicator Service Manual...
Page 154: ...154 T405S Indicator Service Manual...
Page 155: ......
Page 156: ...B TEK Scale LLC 1510 Metric Ave SW Canton OH 44706 3088...