T405S Indicator Service Manual
69
6.5
Ports
If
Type
is set to
Clnt2 x
This stands for Client 2. The socket opens and closes
every time before and after sending data.
Type x
,
Port x
,
Host x
, and
Tmout x
are displayed. These are explained
above.
6.5.3 Protocol
ProtcL
Use this to configure up to 10 communication protocols. These protocols
define the information sent to and from specific types of communication
equipment.
For each protocol you set the type, the binding and the attributes,
described below.
Type x
-
Choose the method to send and receive data for Protocol x,
where x = 1 to 10.
nonE
Choose this to disable the selected protocol.
Choose this when you want to press the
key or when using Autoprint to send the data
through the selected binding (Port) using the
associated attributes.
Some port numbers are reserved. Contact your IT department for the correct port
number to use.
Prot 1
Prot 10
Type x
Bind x
Attr x
Prnft
Inhib
Rate
1
40
Off
On
Enq
Address
1
2
5
10 20
Port 1
Port 2
Enet 1
File
ENQ
SMA
B-Cast
PLC
FTP
Prtr 1
Prtr 2
Prtr.USB
Enet 5
None
RD
NCI
Enq 2
Reports
BSQ
Intprt
Option
50 80 100
Scale
Protcl
Key in
0-254
(0 def.)
Key in
0-254
(5 def.)
Key in
0-254
(5 def.)
Only appears
if configured
in Ztools
Select
scale
number
Which Attributes and Bindings apply to the selected Type will be described in the
detailed descriptions below.
Summary of Contents for T405S
Page 1: ...AWT35 501638 Issue AA T405S Weight Indicator Service Manual...
Page 2: ...BTEK_T405S_s_en_501638 book...
Page 8: ...8 T405S Indicator Service Manual...
Page 134: ...134 T405S Indicator Service Manual...
Page 147: ...T405S Indicator Service Manual 147 13 7 Outline dimensions...
Page 150: ...150 T405S Indicator Service Manual...
Page 154: ...154 T405S Indicator Service Manual...
Page 155: ......
Page 156: ...B TEK Scale LLC 1510 Metric Ave SW Canton OH 44706 3088...