72
T405S Indicator Service Manual
6
ADMIN level menus
6.5.4 P.F.Edit
P.F.Edit
This stands for print format editor. Please refer to the section
for the procedures to edit or create print formats.
6.5.5 PLC
nEt 1
This is an item for future expansion to multiple networks. This is not
available on this indicator. Continue to the next item.
tYPE
In this item you can choose;
EnEt iP
or
M tcP
. Choose
EnEt
iP
to enable Ethernet IP. Choose
M tcP
to enable Modbus-
TCP.
EndiAn
Choices for this item are
big
or
LittLE
. Default is
LittLE
.
Create Print
Formats
P.F. Edit
PrnFt1
PrnFt40
If connecting to a PLC, the E-NET configuration menu does not require any special
configuration and the Client/Server selection for TYPE setting must be set to OFF.
PLC
Type
Endian
Trade
In
Out
Enet IP
M TCP
Little
Big
Off
On
In/Out 1
Type
Value
Sint8
Uint8
Sint16
Uint16
Sint32
Uint32
Float
In/Out 16
Net 1
String
Length
Scale
Only
String
type
Set string
length (0-128)
See Network
Token list
Set scale
number
ENDIAN:
Big = Most Significant Byte, Least Significant Byte
Little = Least Significant Byte, Most Significant Byte
Summary of Contents for T405S
Page 1: ...AWT35 501638 Issue AA T405S Weight Indicator Service Manual...
Page 2: ...BTEK_T405S_s_en_501638 book...
Page 8: ...8 T405S Indicator Service Manual...
Page 134: ...134 T405S Indicator Service Manual...
Page 147: ...T405S Indicator Service Manual 147 13 7 Outline dimensions...
Page 150: ...150 T405S Indicator Service Manual...
Page 154: ...154 T405S Indicator Service Manual...
Page 155: ......
Page 156: ...B TEK Scale LLC 1510 Metric Ave SW Canton OH 44706 3088...