T405S Indicator Service Manual
41
6.2
Calibration Procedure
Pt2-Pt9
The points are numbered 2 through 9 because, internally, the
zero reference point is point 1 and the span point is point 10.
Follow the same steps as described in the SPAN procedure to
enter each linearity point.
6.2.5 Input procedure
ZEro
Use this to enter a value for the zero point.
CountS
Use this to enter a zero point using ADC counts.
mV
Use this to enter a zero point using a mV/V value.
SPan
Accept the flashing displayed span weight (
XXXX
) or key in the span
weight that corresponds with the span ADC or mV/V value.
CountS
Use this to enter a span using ADC counts.
mV
Use this to enter a span using a mV/V value.
Linearity points are cleared if a new span calibration is performed.
Input
Zero
Span
Counts
mV
Counts
mV
XXXX
The BSQ base only allows entry of Counts for Zero or Span points
The Span value is the differential value of the actual Calibration Zero and Span count
(or mV/V) values.
Summary of Contents for T405S
Page 1: ...AWT35 501638 Issue AA T405S Weight Indicator Service Manual...
Page 2: ...BTEK_T405S_s_en_501638 book...
Page 8: ...8 T405S Indicator Service Manual...
Page 134: ...134 T405S Indicator Service Manual...
Page 147: ...T405S Indicator Service Manual 147 13 7 Outline dimensions...
Page 150: ...150 T405S Indicator Service Manual...
Page 154: ...154 T405S Indicator Service Manual...
Page 155: ......
Page 156: ...B TEK Scale LLC 1510 Metric Ave SW Canton OH 44706 3088...