External Sense
External Voltage Sense mode.
Maximum voltage drop allowed at load terminals is 5% of voltage
range.
Note: For 600V models, maximum drop allowed is 3% (20Vdc).
Output Impedance
Programmable in CV mode only.
Resolution:
1 mOhm
Range:
0 mOhm to N mOhm.
N = Voltage FS / 2 * Current FS. Voltage will not be reduced
below 50% of programmed value.
OVP Range:
0 to 110 % of V Range
105 to 115 % of V set
0 to 1.1 x Vmax range value
Transient Response:
A 30% step load will recover to within 2% of original value within 2
msec.
Stability:
+/- 0.05% of set point after 8 Hrs. warm up at fixed line, load and
temp
2.1.5
Ranges, Noise and Ripple
Model
Output
Power
Max.
Volts
Max.
Amps
Max.
Typical
RMS Ripple
Typical
PkPk Ripple
Max
PkPk Ripple
KDC 30-167
KDC 30-333
KDC 30-500
5
10
15
30
30
30
167
333
500
15 mV
15 mV
15 mV
45 mV
45 mV
45 mV
90 mV
90 mV
90 mV
KDC 40-125
KDC 40-250
KDC 40-375
5
10
15
40
40
40
125
250
375
15 mV
15 mV
15 mV
45 mV
45 mV
45 mV
90 mV
90 mV
90 mV
KDC 50-100
KDC 50-200
KDC 50-300
5
10
15
50
50
50
100
200
300
15 mV
15 mV
15 mV
45 mV
45 mV
45 mV
90 mV
90 mV
90 mV
KDC 80-62
KDC 80-125
KDC 80-187
5
10
15
80
80
80
62.5
125
187.5
25 mV
25 mV
25 mV
100 mV
100 mV
100 mV
250 mV
250 mV
250 mV
KDC 100-50
KDC 100-100
KDC 100-150
5
10
15
100
100
100
50
100
150
25 mV
25 mV
25 mV
125 mV
125 mV
125 mV
250 mV
250 mV
250 mV
KDC 150-33
KDC 150-66
KDC 150-100
5
10
15
150
150
150
33.3
66.7
100
25 mV
25 mV
25 mV
150 mV
150 mV
150 mV
300 mV
300 mV
300 mV
KDC 250-17
KDC 250-33
KDC 250-50
5
10
15
250
250
250
20
40
60
100 mV
100 mV
100 mV
200 mV
200 mV
200 mV
400 mV
400 mV
400 mV
KDC 300-17
KDC 300-33
KDC 300-50
5
10
15
300
300
300
17
33.3
50
100 mV
100 mV
100 mV
200 mV
200 mV
200 mV
400 mV
400 mV
400 mV
Summary of Contents for KDC 100-100
Page 2: ......
Page 10: ......
Page 26: ...User Manual 16 KDC Series II DC Power Supplies Figure 3 1 Location of rear panel connectors ...
Page 33: ...User Manual KDC Series II DC Power Supplies 23 Figure 4 3 Front panel layout ...
Page 82: ......
Page 83: ...User Manual KDC Series II DC Power Supplies 73 Figure 6 1 DC Power Supply Block Diagram ...
Page 111: ...User Manual KDC Series II DC Power Supplies 101 Figure 8 2 Replaceable Fuse Locations ...