7 Calibration
Routine calibration should be performed every 12 months. Non-routine calibration is only
required if a related assembly is replaced or if the periodic calibration is unsuccessful.
All routine calibrations can be performed without removing the top cover (Closed case
calibration) through the front panel menus or over the remote control interface.
7.1 Calibration Equipment
Equipment
Description
Digital Multimeter (DMM):
Two (2) HP 34401A
0.1%, 10 miliohm Current Shunt or
0.1%, 1 miliohm Current Shunt:
Isotek Model RUG-Z-R010-0.1 or equivalent or
Isotek Model RUG-Z-R001-0.1 or equivalent
Load Bank:
Various power load resistors will be needed
Model
Voltage Range
Power
max.
Current
max.
Load
KDC30-XXX
30
5 KW
167
0.18
10 KW
333
0.09
15 KW
500
0.06
KDC40-XXX
40
5 KW
125
0.32
10 KW
250
0.16
15 KW
375
0.107
KDC50-XXX
50
5 KW
100
0.5
10 KW
200
0.25
15 KW
300
0.167
KDC80-XXX
80
5 KW
62
1.29
10 KW
125
0.64
15 KW
187
0.428
KDC100-XXX
100
5 KW
50
2.0
10 KW
100
1.0
15 KW
150
0.67
KDC150-XXX
150
5 KW
33
4.545
10 KW
67
2.273
15 KW
100
1.5
KDC250-XXX
250
5 KW
20
12.5
10 KW
40
6.25
15 KW
60
4.167
KDC300-XXX
300
5 KW
17
17.647
10 KW
33
8.823
15 KW
50
6
KDC400-XXX
400
5 KW
12.5
32
10 KW
25
16
15 KW
37.5
10.67
KDC600-XXX
600
5 KW
8.33
72
10 KW
16.67
36
15 KW
25
24
Table 7-1: Load Resistors and Current by model
Summary of Contents for KDC 100-100
Page 2: ......
Page 10: ......
Page 26: ...User Manual 16 KDC Series II DC Power Supplies Figure 3 1 Location of rear panel connectors ...
Page 33: ...User Manual KDC Series II DC Power Supplies 23 Figure 4 3 Front panel layout ...
Page 82: ......
Page 83: ...User Manual KDC Series II DC Power Supplies 73 Figure 6 1 DC Power Supply Block Diagram ...
Page 111: ...User Manual KDC Series II DC Power Supplies 101 Figure 8 2 Replaceable Fuse Locations ...