9.7 Trouble Shooting - IEEE-488 / GPIB
This section provides guidelines for resolving communication problems that may occur when
using the 2001RP with or without the KDCGUI software with a IEEE-488 / GPIB interface.
Symptoms:
1. Unable to verify connection to the power supply through IEEE-488 interface. Time-out occurs
any time an attempt is made to verify connection.
Things to check:
1. Is the interface option set to use the IEEE-488 interface instead of the RS232C interface?
The DIP switch on the right side panel of the power supply is used to select the desired
interface mode. Refer to section 0. If the GPIB option (option -IF) was ordered, it should be
set this way from the factory but may have been changed after shipment).
2. Does the bus address selected in the System, Interface dialog match the address set using
the DIP switches on the 2001RP? The GPIB address selected must match the settings on
the DIP switch. If changes are made to the DIP switch setting, the power supply must be
cycled off and on to have the new settings take effect.
3. Did you connect the cable in correctly and are both ends of the GPIB cable securely tied
down?
Select the GPIB interface and click on the NI-488.2M Settings tab. An Advanced button at the
bottom of the screen will provide access to the Bus Timing parameter. Set this value to 2 sec to
operate with the power supply's GPIB interface.
9.8 Software Registration
Updates of this and other Argantix programs are posted on a regular basis on the Argantix web
site. You can find available programs by selecting the Software, GUI's and Drivers menu. To
gain access to these downloads, you will need to register as a user on the Argantix web site. For
instructions on how to register and obtain the required access level for software and user manual
downloads, visit our web site at
www.argantix.com
Summary of Contents for KDC 100-100
Page 2: ......
Page 10: ......
Page 26: ...User Manual 16 KDC Series II DC Power Supplies Figure 3 1 Location of rear panel connectors ...
Page 33: ...User Manual KDC Series II DC Power Supplies 23 Figure 4 3 Front panel layout ...
Page 82: ......
Page 83: ...User Manual KDC Series II DC Power Supplies 73 Figure 6 1 DC Power Supply Block Diagram ...
Page 111: ...User Manual KDC Series II DC Power Supplies 101 Figure 8 2 Replaceable Fuse Locations ...