User Manual
38
KDC Series II DC Power Supplies
hand knob (VOLTAGE) is used to increment or decrement the list point sequence number (#).
The sequence number can only be increased to the next available empty (new) list point.
To move to the next or previous parameter, use the UP (
) or DOWN (
) cursor keys
The voltage and current slew parameters can be set to their maximum slew rates by turning the
CURRENT knob counter clock-wise past 0. This will cause the slew parameter to 'wrap around' to
its maximum available value. This value is a function of the voltage and current range and is
different for different models.
It is not necessary to use all list points, only as many needed to accomplish the desired output
sequence.
Setting Data Values
Data values can be set for each point in a list. If all data values in a specific list are going to be the
same value (e.g. the current limit parameter is set to the same value for the entire transient
program), only the first data value for that parameter has to be set. Setting only the first data point
will automatically repeat that value for all subsequent points in the transient list.
The programming resolution from the front panel for setting data values for voltage is a function of
the voltage range. The programming resolution from the front panel for setting data values for
current is a function of the current range, which is determined by the voltage and power range.
Table 4-2 shows these values as a function of voltage range and power level.
Voltage Range
Voltage
resolution (V)
Current
Resolution (A)
Current
Resolution (A)
Current
Resolution (A)
5 KW
10 KW
15 KW
30 V
0.01
0.05
0.1
0.2
40 V
0.01
0.05
0.1
0.2
50 V
0.02
0.1
0.1
0.1
80 V
0.02
0.02
0.1
0.1
100 V
0.1
0.02
0.1
0.1
150 V
0.1
0.01
0.02
0.1
250 V
0.1
0.01
0.01
0.02
300 V
0.1
0.01
0.01
0.02
400 V
0.1
0.01
0.01
0.01
600 V
0.2
0.01
0.01
0.01
Table 4-2: Transient programming resolution from front panel.
Summary of Contents for KDC 100-100
Page 2: ......
Page 10: ......
Page 26: ...User Manual 16 KDC Series II DC Power Supplies Figure 3 1 Location of rear panel connectors ...
Page 33: ...User Manual KDC Series II DC Power Supplies 23 Figure 4 3 Front panel layout ...
Page 82: ......
Page 83: ...User Manual KDC Series II DC Power Supplies 73 Figure 6 1 DC Power Supply Block Diagram ...
Page 111: ...User Manual KDC Series II DC Power Supplies 101 Figure 8 2 Replaceable Fuse Locations ...