User Manual
KDC Series II DC Power Supplies
85
5. Increment or decrement the RPV 5V calibration coefficient while monitoring the output
voltage until the measured voltage matches the programmed setting.
Full Scale - External R Input
1. Change the DC supply control mode for VOLT to REF EXTRES from the CONTROL screen.
2. Use a suitable resistor between the Remote R voltage programming input.
3. Monitor the output voltage using an external DMM
4. Select the OUTP CAL, RPV RES screen.
5. Increment or decrement the RPV RES calibration coefficient while monitoring the output
voltage until the measured voltage matches the programmed setting.
7.4.2
Output Current Calibration
Offset
1. Set the load to Constant Current mode.
2. Apply a load that will draw more than 2 % of full-scale current.
3. Set the output voltage high enough to force the DC supply in CC mode.
4. Monitor the output current using an external current shunt and DMM
5. Select the OUTP CAL, CURR OFST screen.
6. Increment or decrement the CURR OFST calibration coefficient while monitoring the output
current until the measured current matches the programmed setting.
Note:
It may be convenient to perform the current measurement offset
calibration at this point as well.
Full Scale - Internal
1. Apply a load that will draw more close to 90 % of full-scale current.
2. Monitor the output current using an external current shunt and DMM
3. Select the OUTP CAL, CURR F/S screen.
4. Increment or decrement the CURR F/S calibration coefficient while monitoring the output
current until the measured current matches the programmed setting.
Note:
It may be convenient to perform the current measurement full-scale
calibration at this point as well.
The following calibrations are required only if the analog programming option is installed
(Option -IF)
Full Scale - External 10 V Input
1. Change the DC supply control mode for CURR to REF EXT10V from the CONTROL screen.
2. Use a small precision DC supply or DC calibrator to apply a 10 V DC input to the RPI input.
3. Monitor the output current using an external current shunt and DMM
4. Select the OUTP CAL, RPI 10V screen.
5. Increment or decrement the RPI 10V calibration coefficient while monitoring the output
current until the measured current matches the programmed setting.
Summary of Contents for KDC 100-100
Page 2: ......
Page 10: ......
Page 26: ...User Manual 16 KDC Series II DC Power Supplies Figure 3 1 Location of rear panel connectors ...
Page 33: ...User Manual KDC Series II DC Power Supplies 23 Figure 4 3 Front panel layout ...
Page 82: ......
Page 83: ...User Manual KDC Series II DC Power Supplies 73 Figure 6 1 DC Power Supply Block Diagram ...
Page 111: ...User Manual KDC Series II DC Power Supplies 101 Figure 8 2 Replaceable Fuse Locations ...