User Manual
KDC Series II DC Power Supplies
53
4.5 Setting the Power-on Initialization Values
The power supply is shipped with default factory settings when the unit is powered up. The factory
settings are:
Parameter
Factory default setting
Voltage
0.0 Volt
Current limit
Maximum available current.
Display mode
Current measurement
Output state
OFF
Local / Remote State
Local. Front panel unlocked.
Table 4-5: Factory Default Power on Settings
It is possible to change the power on initialization values in one of two ways:
1. Using the RS232 or optional IEEE-488 interface and the supplied KDCGUI program.
2. Using the front panel.
To change the power on initialization values from the front panel, proceed as follows:
1. Set the unit up in the desired way from the front. (Voltage, current limit, output state, display
mode).
2. Press the MENU button to select the top-level menu and scroll down to the REGISTERS
entry. Press ENTER to select the configuration menu.
3. In the REGISTERS screen, select any number from 0 through 7 you want to use to save this
setup. Press ENTER to save the present setup and remember the NVM register number
used.
4. Press the MENU button to select the top-level menu and scroll down to the CONFIGATION
entry. Press ENTER to select the configuration menu.
5. In the configuration screen, scroll down to the POWER ON REG entry. With the cursor on the
POWER ON REG field, use the knob to scroll to the NVM number used in step 3.
6. Press ENTER to make this register the new power-on default.
7. This procedure can be repeated as often as needed by the user.
Summary of Contents for KDC 100-100
Page 2: ......
Page 10: ......
Page 26: ...User Manual 16 KDC Series II DC Power Supplies Figure 3 1 Location of rear panel connectors ...
Page 33: ...User Manual KDC Series II DC Power Supplies 23 Figure 4 3 Front panel layout ...
Page 82: ......
Page 83: ...User Manual KDC Series II DC Power Supplies 73 Figure 6 1 DC Power Supply Block Diagram ...
Page 111: ...User Manual KDC Series II DC Power Supplies 101 Figure 8 2 Replaceable Fuse Locations ...