7.2 Calibration Password
All Calibrations on the KDC series can be made from the front panel or over the bus. The
Calibration screens are protected by a password to prevent unauthorized calibration however.
When selecting either the MEAS CAL or OUTPUT CAL screens, the operator will be prompted
for a calibration password.
The calibration password is equivalent to the maximum voltage range value. Thus, for a KDC
100-50 model, the password is "100". The password can be entered by dialing it in with one of
the rotary knobs on the front panel or sent over the bus. See programming manual (P/N 5005-
961) for details on bus operation.
Once entered, the calibration system will be unlocked until power on the DC supply is cycled.