MM-L300A
6. How to Operate the PC Software
6-29
(6) Editing the upper/lower limit judgment value [Ratio setting]
Set the upper/lower limit judgment value for determining whether the obtained
waveform is normal or abnormal.
Adjust and edit the upper/lower limit of the reference waveform data created on the
Basis Waveform Calculation Window.
* When the waveform data does not exist before the waveform data is obtained,
default settings are as follows: average mode, average waveform intensity 32767,
upper limit 50%, and lower limit -50%.
1) Set the setting mode to [Ratio setting].
2) Specify the data section to set the upper/lower limit judgment value by
operating scroll bars of [Edit Start Point] and [Edit End Point], or [+]/[-] button.
For [Edit End Point], set the same value as that of [Edit Start Point] or larger.
If the setting range conditions above is not satisfied, a message appears when a
condition is changed or added.
Click the [OK] button to return.
Set upper/lower limit
judgment
value of this section.