7. Basic Operation
7-2
5) Select the condition number to set conditions by clicking from the Condition
Configuration List and click the [Select] button or double-click the condition
number to set.
6) Set the measurement time of waveform on the Condition Configuration
Window.
For the measurement time, set a time longer than the desired laser irradiation
time.
The waveform trends of each sensor unit are shown below.
SU-N300A
:
When the time at least 100 times as long as the laser irradiation time is set, the
sampling cycle becomes large and waveforms obtained from the
MM-L300A
become likely to be varied. (Refer to examples on the next page.)
(Example) Waveform resolution when the measurement time is set to 100 times
Laser irradiation time
Measurement time
Sampling cycle *
0.1 ms
10 ms
4
μ
s
1 ms
100 ms
34
μ
s
10 ms
1 s
334
μ
s
100 ms
10 s
3.334 ms
1 s
100 s
33.334 ms
*
When the measurement time is input, the sampling cycle is calculated
automatically and displayed in the item of the sampling cycle under the
measurement time.