You can also abort the test at any time by pressing the
Abort
button.
Once the measurement has finished, the jitter composition shown in the
Jitter Setup
window is restored.
When the measurement is either completed or aborted, you can obtain
a list of measured points by clicking the following tab which appears
once the measurement is stopped.
Clicking on the above tab you get the list of measured points.
Selecting a point, and clicking on 'Set', updates the Jitter Setup page
with the jitter setting of that point.
Generate HTML Report
The measurement results can be saved as HTML file. To generate the
HTML report:
1
Click the 'Jitter' tab on the toolbar. The following drop down list will
appear.
2
Select 'Generate HTML report'.
3
Select the folder where you want to save the report, name the folder,
and save it.
Jitter Tolerance Characterization - Reference
Jitter tolerance characterization determines the actual jitter levels
where the device under test can no longer maintain a desired BER.
7
Jitter Tolerance Tests
348
Agilent J-BERT N4903 High-Performance Serial BERT
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...