To become active, jitter sources must be switched on (by enabling the
corresponding checkbox) and must at least have an
Amplitude
greater
than zero.
The
Overview
column consists of buttons that can be pressed to change
the parameters of each jitter type independently from others.
The
Overview
button can be used to return after changing parameters.
Random Jitter Parameters
The random jitter
Amplitude
must be entered as an rms (root mean
square) value. The rms value corresponds to the sigma factor
σ
of the
bell-shaped Gauss distribution curve (see also
).
7
Jitter Tolerance Tests
330
Agilent J-BERT N4903 High-Performance Serial BERT
Amplitude