Set Mask Scaling
With mask scaling the mask can be manually enlarged, shrunk, and
moved around. With the logic levels track selected, the mask regions are
bound together, and move together with a fixed amplitude. Position
(pos) gives the timing displacement, and dT gives the delta time offset.
Set Mask Run Control:
The Mask Run Control aligns the mask, switches it on, starts displaying
the number of waveforms, and the number of mask violations per mask
region, and the worst BER per Mask region.
Exit stops the computation, hides the mask, and the violations.
Automated Eye Parameter Measurement
Table 26
S.No
Parameter
Description
1
Rise time
Measures the mean
transition-time taken by
the data on the rising edge
of the eye diagram. The
data crosses three
thresholds: lower,
crossover point, upper and
the eye transition.
Note: Rise time is
dependent on the
Transition Time of the
View Tab. The two options
are 10/90 and 20/80.
5
Advanced Analysis
272
Agilent J-BERT N4903 High-Performance Serial BERT
Mask Scaling
Mask Run Control
Automated Eye Parameter
Measurement
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...