Inputs and Outputs - Procedures
To select the termination for the error detector:
N O T E
Prerequisite: You must know the termination voltage of the data signal that your
DUT sends to the error detector.
1
In the
ED Setup
menu, press
Sampling Point Setup
to access the
Sampling
Point Setup
window.
2
Click the
Edit
button. This opens the
ED Input Setup
dialog.
3
In the
Termination
field, enter the termination voltage that is
appropriate for the incoming data signal.
You can use the front panel knob or the numeric keypad to enter the
value.
C A U T I O N
Selecting the wrong termination may damage your device.
4
Select the
Data Inverted
checkbox if your device inverts data.
You can now physically connect the DUT to the error detector.
Data Input Setup - Reference
The
ED Input Setup
dialog is accessed from the
Sampling Point Setup
window. It is used to set up the error detector's data input port:
Input
The selection in this list defines how the signals arriving at the Data In
and Data In connectors are interpreted. The following options are
available:
•
Differential
4
Setting up the Error Detector
114
Agilent J-BERT N4903 High-Performance Serial BERT
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...