The default is 0.01 UI, that means, 100 points per unit interval will
be measured.
•
Edge Resolution Optimization
Turns the resolution optimization on or off. If this option is enabled,
the Serial BERT intelligently sets a resolution so that there are more
sampling points at the edges of the eye. This can greatly improve the
results without dramatically increasing the duration of the test.
Set the criteria for the sample threshold:
•
Resolution
Specifies how many measurement points are taken within the
sample voltage range. Enter the value in mV.
•
Low Level
Specifies the lower limit of the measurement voltage range. Enter
the value in mV. This value should be slightly lower than the lowest
expected signal voltage.
•
High Level
Specifies the upper limit of the measurement voltage range. Enter
the value in mV. This value should be slightly higher than the highest
expected signal voltage.
Pass/Fail Tab
The
Pass/Fail
tab of the
Properties
dialog allows you to specify the criteria
to decide whether the DUT passes or fails the test. You can change pass/
fail criteria without rerunning the test. The software only uses the
criteria to rate the results of a measurement.
N O T E
The pass/fail criteria do not control measurement execution. The measurement run
will be completed even if the measurement fails one or more of the criteria.
You can set several pass/fail limits. For an explanation of the
Eye Opening
Pass/Fail
“Measurement Parameters” on page 235
.
5
Advanced Analysis
230
Agilent J-BERT N4903 High-Performance Serial BERT
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...