Problems with the N4916A
Problems with the N4916A - Concepts
When you have opened the
PG Output Setup
window and press
De-
Emphasis
, the N4916A Connection window appears.
Cannot-Connect Problems
An error message may appear when you now press
Connect
.
If this happens, ensure that the N4916A has been correctly installed (see
“Installing the N4916A - Procedure” on page 407
).
Check the Power Supply
Ensure that the power cord is connected.
The green LED close to the power switch must be illuminated when the
unit is turned on.
Check the USB Communication
Ensure that the USB cable is connected between the USB port at the
rear of the N4916A and one of the USB ports at the rear of the Serial
BERT.
1
Open the
Utility
menu and click
Minimize GUI
to gain access to the
Windows operating system.
2
In the Windows task bar, click the Agilent IO Control icon.
8
Solving Problems
386
Agilent J-BERT N4903 High-Performance Serial BERT