Set the criteria how the sampling points of the eye are displayed:
•
Timing Unit
Choose between
Unit Interval
and
Seconds
to set the timing values for
the measurement points on the
Parameters
tab. This setting does not
affect the display of results.
•
Threshold Type
Choose between
Absolute, Offset,
or
Percentage
to set the thresholds for
the measurement points:
–
Absolute
You specify the thresholds for the measurement points as
absolute voltages.
–
Offset
You specify the thresholds for the measurement points relative
to the threshold voltage of the actual sampling point.
–
Percentage
You specify the thresholds for the measurement points as a
percentage of the current eye opening voltage of the error
detector.
Depending on the quality and characteristics of the eye opening,
the resulting 50% threshold may deviate from the decision
threshold defined in the loaded setting.
•
Number of Valid Points
Change the
Number of Valid Points
, if you wish to use less or more
measurement points for the measurement.
Advanced Analysis
5
Agilent J-BERT N4903 High-Performance Serial BERT
247
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...