Setting the Frequency Range
Choose
Auto
or
Manual
mode.
In
Auto
mode, the jitter frequencies are automatically calculated:
1
Enter the
Start Frequency.
2
Enter the
Stop Frequency
.
Start and stop frequency define the measurement's frequency range.
3
Set the
Number of Steps
between start and stop.
The
Number of Steps
yields a set of logarithmically equidistant
frequencies.
In
Manual
mode, the automatically calculated jitter frequencies can be
edited. You can also enter your own comma-separated list of
frequencies.
The
Show Compliance Curve
checkbox enables the compliance curve on
the Jitter Tolerance Characterization graph. This feature is available
only if the compliance jitter measurement option is installed. The
standard is taken from the compliance measurement.
7
Jitter Tolerance Tests
344
Agilent J-BERT N4903 High-Performance Serial BERT
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...