Sinusoidal Interference
To enable the addition of sinusoidal interference, press the
corresponding checkbox.
Set the parameters to desired values:
The peak-to-peak amplitude of sinusoidal interference must not exceed
a certain level (see the technical specifications).
The frequency range of sinusoidal interference is limited (see the
technical specifications).
One of four modes can be chosen for sinusoidal interference:
•
Common
: Requires that a differential data signal is generated and
connected to P1/P1 or P2/P2 of the Interference Channel.
In common mode, both components of the differential data signal
are amplitude-modulated with the same sinewave. This adds
common mode noise.
•
Differential
: Requires that a differential data signal is generated and
connected to P1/P1 or P2/P2 of the Interference Channel.
In differential mode, both components of the differential data signal
are amplitude-modulated, one with the normal sinewave, the other
with the inverted sinewave. This adds differential noise.
•
Single-ended normal
: Modulates the data signal that is connected to P1
or P2 of the Interference Channel.
•
Single-ended complement
: Modulates the data signal that is connected
to P1 or P2 of the Interference Channel.
Jitter Tolerance Characterization
Jitter Tolerance Characterization - Concepts
During a jitter tolerance characterization test, jitter is applied
sequentially at a number of frequency points in the frequency range of
interest.
7
Jitter Tolerance Tests
338
Agilent J-BERT N4903 High-Performance Serial BERT
Amplitude
Frequency
Mode
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...