2
Use the file dialog box to locate the desired .jcp file and press
Open
to load the measurement.
Accessing Comment and Condition Text
To access the comment or condition text of a measurement:
•
Double-click
Modify Comment
below the graphical display.
This opens the
Jitter Measurement Comment
dialog box.
You can enter a comment. The condition is automatically added by the
system.
You can also read the full comment or condition in case only part of the
text is displayed in the result table.
Jitter Tolerance Compliance - Reference
Jitter Tolerance Compliance determines whether a device under test
maintains a desired BER at defined jitter levels. The jitter levels can be
derived from predefined standards.
Standard Specification
The drop-down list contains all predefined standards, including user-
defined standards. The shape of the selected standard is shown in the
preview area. The frequency range and the jitter amplitude range of the
standard is given in the preview. A description is displayed at the
bottom of the
Standard
tab.
The last entry in the list is
Browse…
. Select this entry to load a file that
contains a user-defined standard. See
“Setting the Standard” on page
To aggravate test conditions, the standard's test jitter amplitude can be
increased by
Margin
percent. The display shows a black line for the
standard and a green line for the new jitter amplitudes.
Select the
Auto Frequencies
checkbox to set frequencies that are
recommended for the selected standard. This list contains, for example,
the edge frequencies of the standard.
7
Jitter Tolerance Tests
368
Agilent J-BERT N4903 High-Performance Serial BERT
Standard Selection
Margin
Auto Frequencies / Number of Steps
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...