Installing Hardware Options
Installing Hardware Options - Procedures
The Serial BERT has a slot for installing optional hardware.
At the present time, the J20 Interference Channel is available. This
module can be used for adding intersymbol interference and sinusoidal
interference to the signal generated by the Pattern Generator.
Installing the J20 Option
To install the J20 Interference Channel:
1
Perform the regular shutdown procedure provided by the operating
system.
2
Disconnect the instrument from mains.
C A U T I O N
Never plug in or remove a module while the instrument is connected to mains power.
3
Unscrew and remove the slot cover.
4
Insert the module into the slot.
5
Press the lever until the module rests firmly in place, the front plate
aligned with the Error Detector's front panel.
6
Insert and tighten the fixing screw.
7
Power up the instrument. The software will automatically recognize
the new component and provide the Interference Channel Icon and
according setup functions in the
Jitter
window.
9
Customizing the Instrument
406
Agilent J-BERT N4903 High-Performance Serial BERT
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...