The oscilloscope should show the following picture:
If you see only a straight line, ensure that the pattern generator is not
disabled and that it is set up to generate a pattern (e.g. PRBS).
Check the Output of the Pattern Generator with the Error Detector
If you have connected the N4916A to a Serial BERT you can of course
check the output of the pattern generator by means of the error detector.
1
Set up the pattern generator as described above: Set the Data Offset
to 0 V and the Amplitude to 1 V.
2
Set up the error detector as described above.
3
Connect the Data Output of the Serial BERT to the Data Input of the
error detector.
4
Ensure that the error detector follows the generated pattern.
– Press the
Auto Align
button.
– From the
Analysis
panel select the
Eye Diagram
page.
– Press the
Start
button.
8
Solving Problems
392
Agilent J-BERT N4903 High-Performance Serial BERT
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...